16.2.2011
München |
|
|
|
|
|
|
BaSyTec Battery-Test-Systems
HPS Series
Technical data
|
Current range
|
up to 500 A (parallel connection of channels possible)
|
|
Voltage range
|
up to 60 V
|
|
Power
|
100 W to 10 000 W (for one channel)
Parallel connection increases the power
|
|
Measurement resolution
|
0.003 % (16Bit)
|
|
Measurement accuracy
|
0.05 %
|
|
Measurement method
|
synchronic measurement of voltage, current and aux channel
|
|
Controlable values
|
Current, voltage: Analogue control
Power, resistance: Digital control
combinations: Digital control
|
|
Calculated values
|
power, charge, energy, deviations, others
|
|
Additional input channels
(optional)
|
Up to four for each test channel (i.e. Pt 100 or voltage input (100 mV to 60 V)
|
|
Control input and output
|
Up to seven digital input/output per channel on board
Up to eight relais outputs per channel (with additional relays board)
|
|
Rise time
|
1 ms/A (minimum 30 ms)
|
|
Minimum pulse duration
|
1 ms
|
|
Interface
|
RS 232 and ethernet
|
|
Software protocol
|
Basytec protocol, see BaSyTest Software
|
|
Other options
|
- Battery holders
- Calibration device
- Interface for battery management systems
- Control of external charger / load
- Battery safety device testing extension
- Interface for other power amplifiers
- Interface for climate chambers
- Client server database
|
Last modified: Thursday, January 12, 2006
|
|